Energy Science / BioLogic / Battery Cycling, Instruments

MPG-200 battery cycler series - BioLogic


Discover the MPG-200 Series: A New Era in Battery Testing

The BioLogic MPG-200 series represents a paradigm shift in battery testing technology, combining the precision of high-end potentiostats/galvanostats with the simplicity and efficiency of a battery cycler. Designed to meet the rigorous demands of the most challenging battery testing requirements, the MPG-200 series is equipped with state-of-the-art features to ensure unparalleled performance.

Users can select from 16-channel configurations capable of delivering ±100 mA per channel, or the eight-channel setup offering up to ±5 A per channel, making the system versatile for a wide range of cell types and testing scenarios. The MPG-200 series goes beyond standard testing capabilities by incorporating Electrochemical Impedance Spectroscopy (EIS), available as an optional feature, allowing simultaneous EIS testing across all channels to capture detailed battery behavior from 100 kHz down to 10 µHz.

Each channel in the MPG-200 series can be enhanced with a Pt-100 probe for precise temperature monitoring, alongside seamless integration with external devices through both analog and digital I/O. Designed for scalability, units can be easily stacked in a five-position rack, offering tailor-made solutions for any testing requirement. Whether for in-house R&D or academic research, the MPG-200 series stands as the epitome of battery testing excellence, perfectly suited for energy applications and supported by EC-Lab® software, the benchmark in electrochemical testing software.

Technical information

MPG-2

MPG-205

Channel number168
Cell connection2, 3, 4 or 5 terminal leads2 or 4 terminal leads
Voltage
Range± 10 V-2 V: 9 V
Control resolution200 µV down to 5 µV
Measurement resolution0.004 % of FSR
Accuracy± 0.1 % of control ± 0.01% of FSR
Current
Max (continuous) per channel± 100 mA± 5 A
Ranges5: 100 mA down to 10 µA7: 5 A down to 10 µA
Control resolution0.004 % of FSR /0.8 nA
Measurement resolution0.004 % of FSR
Accuracy± 0.1 % of control ± 0.01 % of FSR
EIS
Built-inOptional: simultaneously on every channel
Range10 µHz to 100 kHz
Measurement
Acquisition time200 µs
Time base200 µs
Auxiliary inputs/outputs
Monitor ouputE and I monitorsI monitor
2 analog inputsAutomatic ± 2.5 V, ± 5 V, ± 10 V ranges – 16 bits resolution
1 analog output± 10 V range, 16 bits resolution
2 digital inputsTTL level trigger input
1 digital ouputTTL level trigger ouput
General
Dimensions (H x W x D)260 x 495 x 465 mm254 x 494 x 454 mm
Weight17 kg25 kg
Power consumption350 W860 W

Documents & videos